ENERGY OUTPUT MEASUREMENT
HIGH RANGE:
Voltage: < 5000 V
Maximum current: 120 A
Maximum Energy: 1000 J
Accuracy: ±2 % of reading for > 100 J
±2 J of reading for < 100 J
Trigger level: 100 volts
Playback amplitude: 1 mV/1000 V Lead I
Test pulse: 125 J ±20%
LOW RANGE:
Voltage: < 1000 V
Maximum current: 24 A
Maximum Energy: 50 J
Accuracy: ±2 % of reading for > 20 J
±2 J of reading for < 20 J
Trigger level: 20 volts
Playback amplitude: 1 mV/200 V Lead I
Test pulse: 5 J ±20%
Load resistance: 50 ohms ±1 %;
non-inductive (< 1 µH)
Display resolution: 0.1 J
Measurement time Window: 100 ms
Absolute max. peak Voltage: 6000 V
Pulse width: 100 ms
Cardioversion: Measured time delay ±2 ms
Biphasic Measurement
OSCILLOSCOPE OUTPUT:
High Range: 1000:1 amplitude attenuation
Low Range: 200:1 amplitude attenuation
WAVEFORM STORAGE AND PLAYBACK:
Discharge can be viewed via ECG outputs and
paddles
Output: 200:1 Time Base Expansion
SYNC TIME MEASUREMENTS:
Timing window: Starts 40 ms before each R
wave peak
Test waveforms: All Waveform simulations
available
Delay time accuracy: ±1 ms
CHARGE TIME MEASUREMENT:
From 0.1 to second to 99.9 seconds
ECG WAVE
ECG GENERAL:
Lead configuration: 12-lead simulation.
RL, RA, LA, LL, V1-6
Output Impedance:
Limb leads: 1000 ohms to RL
V Leads: 1000 ohms to RL
High level Output: 1.0 V/mV of low level (lead II)
MANUAL ECG PERFORMANCE TEST:
DC Pulse: 4.0 seconds 1.0 mV
Square wave: 2 Hz 1.0 mV p-p
Triangle wave: 2 Hz 1.0 mV
Sine waves: 0.1, 0.2, 0.5, 10, 40, 50, 60, 100Hz
Amplitudes: 0.5, 1.0, 1.5, 2.0 mV (Lead II)
Accuracy: ±5 % (Lead II 1.0 mV)
ECG PERFORMANCE TEST:
Gain/Damping: 2 Hz square wave
Frequency response:
Low Frequency: 4 second DC pulse
Band Pass: 10 Hz sine
Monitor: -3dB point: 40 Hz sine
Power line notch filter: 50 Hz sine
Linearity: 2 Hz triangle wave
NORMAL SINUS:
Rates: 30, 60, 80, 120, 180, 240 & 300 BPM
Accuracy: ±1 % of selection
Amplitudes: 0.5, 1.0, 1.5, 2.0 mV (Lead II)
Accuracy: ±5 % (Lead II 1.0 mV)
Automatic ECG rate Test
ARRHYTHMIA SELECTIONS:
vfib – Ventricular Fibrillation
afib – Atrial Fibrillation
blk II – Second degree A-V block
RBBB – Right Bundle Branch Block
PAC – Premature Atrial Contraction
PVC_E – Early PVC
PVC_STD – PVC
PVCRonT – R on T PVC
mfPVC – Multifocal PVC
bigeminy – Bigeminy
run5PVC – Run of 5 PVCs
vtach – Ventricular Tachycardia
SHOCK ADVISORY ALGORITHM TEST:
8 ECG signals are available to test the advisory
algorithm of Automatic defibrillators:
Asystole
CVF: Coarse Ventricular Fibrillation
FVF: Fine Ventricular Fibrillation
MVT 140: Multifocal Ventricular Tachycardia @ 140 BPM
MVT 160: Multifocal Ventricular Tachycardia @ 160 BPM
PVT 140 : Polymorphic Ventricular Tachycardia @ 140 BPM
PVT 160 : Polymorphic Ventricular Tachycardia @ 160 BPM
SVTa 90 : Supra Ventricular Tachycardia @ 90 BPM